UNITY - Integrated Electrolyte System
 

UNITYTM

Integrated Electrolyte System

The Diamond Diagnostics Unity Integrated Electrolyte System is a fast and compact ISE module measuring Na+, K+ and Cl- within Serum, Plasma and Diluted Urine utilizing Ion Selective Electrode (ISE) technology. The Diamond Unity is uniquely designed with a powerful microprocessor and integrated electronics allowing for easy integration into any size Chemistry Analyzer platform.

Benefits:

  • Compact Size
  • Flexible Reagent System
  • Long Life Reference
  • Accessible Service Area
  • Fully Visible Pathway
  • Fully Shielded Measurement Chamber
  • Built-In Sample Cup with Overflow Management
  • Positive Engagement Latching System
  • Horizontal Electrode Orientation
  • Precise Stepper Motor Fluidic Control
  UNITY

Watch the Video

SPECIFICATIONS:

Sample
Serum or Diluted Urine

Sample Size
60 - 100 μl Serum
140 -150 μl Urine

Throughput
100 samples/ hour Serum
50 samples/ hour Urine

Communication
RS232 Serial Communication
Programmable Baud Rate

Data Storage
2000 Patient results
500 QC results/ per level (3)
100 Calibrations

Power
12VDC, 2.0A

 

Module Size
92 mm (H) x 134 mm (W) x 46 mm (D)
3.70” (H) x 5.27” (W) x 1.81” (D)
Customized

Reproducibility
Typical Within Run (n=30)
Serum
    Na+: CV ≤ 1% (120-160mmol/L)
    K+ : CV ≤ 2% (2.8-6 mmol/L)
    Cl- : CV ≤ 2% (80-120 mmol/L)
Urine
    Na+: CV ≤ 5% (60-280 mmol/L)
    K+ : CV ≤ 5% (30-170 mmol/L)
    Cl- : CV ≤ 5% (80-230 mmol/L)

Analysis Time
30 seconds Serum (including 1-point calibration)
60 seconds Urine (including 1-point calibration)


Test Ranges

Analyte Units Range Resolution
(Serum)
Na+ mmol/L 50 - 200 0.1
K+ mmol/L 1.50 - 10.00 0.01
Cl- mmol/L 50 - 200 0.1
(Diluted Urine)
Na+ mmol/L 20 - 500 1
K+ mmol/L 5 - 300 1
Cl- mmol/L 10 - 500 1