ProLyte Electrolyte ISE Analyzer
 

ProLyte®

The World’s Most Popular Electrolyte Analyzer

The Diamond Diagnostics ProLyte Electrolyte Analyzer is a completely automated system measuring Na+, K+, Cl- and Li+ in whole blood, serum, plasma and urine utilizing Ion Selective Electrode (ISE) technology. The Diamond ProLyte is uniquely designed with a powerful microprocessor allowing for easy interface with external LIS systems, operation with a full keyboard and bar code scanner and storage and retrieval of over 1000 patient results. The system currently operates in 11 languages.

Buy the Diamond ProLyte Electrolyte Analyzer online.

 

ProLyte

Watch the Video

SPECIFICATIONS:

Sample
Whole Blood, Serum, Plasma, or Urine

Sample Size
100 μl Whole Blood, Serum or Plasma
400 μl diluted (1:10) Urine

Number of Parameters
3 (Simultaneously)

Detection Range
Blood Na+: 45-205 mmol/L
K+ : 1.5-11 mmol/L
Cl- : 45-205 mmol/L
Li+ : 0.15-5.0 mmol/L
Urine Na+: 30-1020 mmol/L
K+ : 20-505 mmol/L
Cl- : 25-505 mmol/L

Reproducibility
Typical Within Run (n=30)

Blood, Serum, Plasma
Na+: CV ≤ 1% (115-160mmol/L)
K+ : CV ≤ 2% (2.5-6 mmol/L)
Cl- : CV ≤ 2% (65-120 mmol/L)
Li+ : SD ≤ 0.03 (0.25-2.0 mmol/L)

Urine
Na+: CV≤ 5% (60-280 mmol/L)
K+ : CV ≤ 5% (30-170 mmol/L)
Cl- : CV ≤ 5% (80-230 mmol/L)

Display Resolution
Na+ : 1 mmol/L
K+ : 0.1 mmol/L
Cl- : 1 mmol/L
Li+ : 0.01 mmol/L

 

Analysis Time
57 sec (Blood)
93 sec (Urine)

Data Storage
1000 Patient Results
500 Level 1 QC Results
500 Level 2 QC Results
500 Level 3 QC Results

Calibration
Automatic or On Demand
1 Point after every Measurement

User Input
Keypad, External Keyboard,
External Barcode Scanner (optional)

Output
Graphic backlit display
40 column printer
AutoSampler port*

Ambient Conditions
15-32°C, Room temperature (60-90°F)
<85% humidity

Languages
English, Chinese, French, German, Italian, Korean, Japanese, Portuguese, Russian, Spanish, Polish

Power
100-240 ~ VAC 50/60 Hz, 1.6 A

Size & Weight
9.5” (24cm) W x 16.5” (42cm) H x
8.0” (20cm) D, 10.5 lbs (4.5 Kg)